Cleared at 93%. Deployed at 64%. Clinical AI drifts
Across real-world scans, the pulmonary embolism triage tool caught 64.1% of cases against a 93% cleared figure, with sensitivity falling 15% a year at one site.
- Deployed sensitivity of the Aidoc BriefCase PE triage tool was 64.1%, against 93% sensitivity reported in the FDA 510(k) summary. Self-reported m:…2203
- Sensitivity was 88.6% for central PE, 63.1% for lobar/segmental PE, and 33.3% for subsegmental PE. Self-reported m:…2203
- Sensitivity differed by 13 percentage points between the two highest-volume sites (68.7% vs 55.7%), and at one site sensitivity fell 15% per year. Self-reported m:…2203
- AToP's Transformer model achieved an AUROC of 0.77 and PR-AUC of 0.34 for 30-day unplanned readmission prediction. Self-reported m:…7967
The gap
An FDA-cleared AI triage tool for pulmonary embolism looked strong in its clearance paperwork and much weaker in the ward. In a preprint posted on medRxiv, researchers report that the deployed sensitivity of the Aidoc BriefCase PE triage tool was 64.1%, against 93% sensitivity reported in the FDA 510(k) summary. The deployed figure comes with a 95% confidence interval of 61.5-66.6% — well below the cleared benchmark, not a borderline miss. m:…2203
The gap is the story. Clearance measures a model under controlled conditions; deployment measures it against real scans, real sites and real drift. The study is a post-market check, and it finds the tool's real-world performance significantly lower than the number on its clearance. m:…2203
Where it misses
Sensitivity varies by where the clot is. The tool caught 88.6% of central pulmonary embolisms, 63.1% of lobar or segmental cases, and 33.3% of subsegmental cases. The tool's sensitivity falls as the embolus gets more peripheral — and subsegmental cases are where it is weakest. m:…2203
Performance also varies by site. Sensitivity differed by 13 percentage points between the two highest-volume sites — 68.7% against 55.7%. The same tool, the same task, and a 13-point gap depending on where it runs. A gap that size between sites running the same tool suggests the difference is not the model alone but the scans, the workflow, or the patients each site sees. m:…2203
The drift
Performance also moves over time. At one site, the false-positive rate fell 40% per year while sensitivity fell 15% per year. The tool was getting quieter and less sensitive at the same time — a pattern the study describes as temporal drift, and one that a one-off clearance check cannot catch. m:…2203
That is the core problem with cleared-but-unmonitored AI: nobody re-runs the clearance test every month, so a slow decline can run for a year before anyone notices. The pattern is only visible because someone kept measuring after deployment. A one-off clearance check cannot catch a decline that unfolds over a year. A tool that is cleared once and never re-checked can drift while everyone assumes the clearance number still holds. m:…2203
The watchers
Other preprints this week build the kind of monitoring the triage study implies is missing. AToP, an explainability framework for Transformer-based predictions from electronic health records, converts per-token Integrated Gradients into structured temporal pattern narratives — visit-anchored stories about why a model made a call, rather than a ranked list of tokens. In its test, the underlying Transformer model trained on MIMIC-IV v3.1 achieved an AUROC of 0.77 and PR-AUC of 0.34 for 30-day unplanned readmission prediction. m:…7967
AToP also surfaces rare, high-attribution markers that standard rankings miss — including psychosis, self-harm overdose, suicidal ideation, and severe MDD — and identifies risk-increasing patterns such as psychiatric sequences and oncologic progressions. MedCheckpoint takes a different angle: it automates postmarketing pharmacovigilance with LLM agents that enforce study-design locks and source-traceable grounding, outperforming baseline coding agents in report quality across five dimensions, with higher run-to-run consistency — and it prevents misleading failures on infeasible comparisons or unmeasurable outcomes. m:…7967 m:…3779
Neither system fixes a drifting triage tool on its own. But together they point at the missing layer in clinical AI: continuous, explainable monitoring after deployment, so that a 64.1% does not pass for a 93% unnoticed. m:…2203
What is still unresolved
- What drives the site-level gap between the two highest-volume sites.
- Whether the drift continues or stabilises over longer deployment.
- Whether explainability frameworks like AToP catch drift in triage tools specifically.
Provenance by Synorb. Every claim links to its source. This story was written by an agent from the Manifests listed below. Read the methodology and limitations, or send a correction.