title: Cleared at 93%. Deployed at 64%. Clinical AI drifts
description: Across real-world scans, the pulmonary embolism triage tool caught 64.1% of cases against a 93% cleared figure, with sensitivity falling 15% a year at one site.
canonical: https://hangingcontext.com/stories/2026-11-13/clinical-ai-drifts-after-deployment/
last-updated: 2026-11-13
url: https://hangingcontext.com/stories/2026-11-13/clinical-ai-drifts-after-deployment/
published: 2026-11-13T00:51:58Z
edition: 2026-11-13T00
newsroom: health-biotech
beat: None
trust: reported
source_count: 3
cited_manifest_ids: ['1794499999680772203', '1794506595633807967', '1794502747804783779']
---

# Cleared at 93%. Deployed at 64%. Clinical AI drifts

*Across real-world scans, the pulmonary embolism triage tool caught 64.1% of cases against a 93% cleared figure, with sensitivity falling 15% a year at one site.*

Reported · 3 Manifests · Health & Biotech · Published 2026-11-13 00:51 UTC

## Key facts

- Deployed sensitivity of the Aidoc BriefCase PE triage tool was 64.1%, against 93% sensitivity reported in the FDA 510(k) summary. (self-reported) [cites: 1794499999680772203]
- Sensitivity was 88.6% for central PE, 63.1% for lobar/segmental PE, and 33.3% for subsegmental PE. (self-reported) [cites: 1794499999680772203]
- Sensitivity differed by 13 percentage points between the two highest-volume sites (68.7% vs 55.7%), and at one site sensitivity fell 15% per year. (self-reported) [cites: 1794499999680772203]
- AToP's Transformer model achieved an AUROC of 0.77 and PR-AUC of 0.34 for 30-day unplanned readmission prediction. (self-reported) [cites: 1794506595633807967]

## The gap

An FDA-cleared AI triage tool for pulmonary embolism looked strong in its clearance paperwork and much weaker in the ward. In a preprint posted on medRxiv, researchers report that the deployed sensitivity of the Aidoc BriefCase PE triage tool was 64.1%, against 93% sensitivity reported in the FDA 510(k) summary. The deployed figure comes with a 95% confidence interval of 61.5-66.6% — well below the cleared benchmark, not a borderline miss. [cites: 1794499999680772203]

The gap is the story. Clearance measures a model under controlled conditions; deployment measures it against real scans, real sites and real drift. The study is a post-market check, and it finds the tool's real-world performance significantly lower than the number on its clearance. [cites: 1794499999680772203]

## Where it misses

Sensitivity varies by where the clot is. The tool caught 88.6% of central pulmonary embolisms, 63.1% of lobar or segmental cases, and 33.3% of subsegmental cases. The tool's sensitivity falls as the embolus gets more peripheral — and subsegmental cases are where it is weakest. [cites: 1794499999680772203]

> **From Synorb** — Claims with a source link. A Manifest holds a Signal of structured claims and a Brief, plus the source it came from. Your agent can check a sentence against it. [Read the guide](https://synorb.com/connect/start)


Performance also varies by site. Sensitivity differed by 13 percentage points between the two highest-volume sites — 68.7% against 55.7%. The same tool, the same task, and a 13-point gap depending on where it runs. A gap that size between sites running the same tool suggests the difference is not the model alone but the scans, the workflow, or the patients each site sees. [cites: 1794499999680772203]

## The drift

Performance also moves over time. At one site, the false-positive rate fell 40% per year while sensitivity fell 15% per year. The tool was getting quieter and less sensitive at the same time — a pattern the study describes as temporal drift, and one that a one-off clearance check cannot catch. [cites: 1794499999680772203]

That is the core problem with cleared-but-unmonitored AI: nobody re-runs the clearance test every month, so a slow decline can run for a year before anyone notices. The pattern is only visible because someone kept measuring after deployment. A one-off clearance check cannot catch a decline that unfolds over a year. A tool that is cleared once and never re-checked can drift while everyone assumes the clearance number still holds. [cites: 1794499999680772203]

## The watchers

Other preprints this week build the kind of monitoring the triage study implies is missing. AToP, an explainability framework for Transformer-based predictions from electronic health records, converts per-token Integrated Gradients into structured temporal pattern narratives — visit-anchored stories about why a model made a call, rather than a ranked list of tokens. In its test, the underlying Transformer model trained on MIMIC-IV v3.1 achieved an AUROC of 0.77 and PR-AUC of 0.34 for 30-day unplanned readmission prediction. [cites: 1794506595633807967]

AToP also surfaces rare, high-attribution markers that standard rankings miss — including psychosis, self-harm overdose, suicidal ideation, and severe MDD — and identifies risk-increasing patterns such as psychiatric sequences and oncologic progressions. MedCheckpoint takes a different angle: it automates postmarketing pharmacovigilance with LLM agents that enforce study-design locks and source-traceable grounding, outperforming baseline coding agents in report quality across five dimensions, with higher run-to-run consistency — and it prevents misleading failures on infeasible comparisons or unmeasurable outcomes. [cites: 1794506595633807967, 1794502747804783779]

Neither system fixes a drifting triage tool on its own. But together they point at the missing layer in clinical AI: continuous, explainable monitoring after deployment, so that a 64.1% does not pass for a 93% unnoticed. [cites: 1794499999680772203]

## What is still unresolved

- What drives the site-level gap between the two highest-volume sites.
- Whether the drift continues or stabilises over longer deployment.
- Whether explainability frameworks like AToP catch drift in triage tools specifically.

## Decision card

- **Useful for:** Clinical AI developers and hospital teams deploying triage tools; Regulators designing post-market monitoring for clinical AI
- **Not for:** Judging an individual patient's scan — this is about population-level tool performance
- **Compare against:** FDA 510(k) clearance benchmarks; Continuous post-market monitoring data

## Provenance ledger

| Manifest id | Title | Source | Claims used | Published |
|---|---|---|---|---|
| `1794499999680772203` | Silent Drift and Site-level Variance in Clinical AI: Real-world performance and subgroup analysis for post-market monitoring of an FDA-cleared artificial intelligence (AI) triage tool for high-risk use case of pulmonary embolism | [medrxiv-diagnostics-imaging-pathology](https://medrxiv.org/content/10.64898/2026.11.10.26367739v1.full) | 1794500128439772548, 1794500128442772549, 1794500128443772550, 1794500128444772551 | 11-12 00:00 |
| `1794506595633807967` | AToP: An Explainability Framework for Transformer-Based EHR Predictions Using Attribution-Guided Temporal Pattern Mining | [medrxiv-health-ai-informatics](https://medrxiv.org/content/10.64898/2026.11.10.26367753v1.full) | 1794506818127808140, 1794506818123808139, 1794506818130808141, 1794506818132808142 | 11-12 00:00 |
| `1794502747804783779` | Automating postmarketing pharmacovigilance with large language model agents | [medrxiv-health-ai-informatics](https://medrxiv.org/content/10.64898/2026.11.10.26367711v1.full) | 1794503069349785571, 1794503069351785572, 1794503069352785573 | 11-12 00:00 |

## Run these Manifests yourself

Every Manifest cited here can be read in full with your own Synorb Key. Check the claims before you rely on them.

1. **Get a Synorb Key.** Click Generate Free Agent Key at https://synorb.com/connect/start, or let your agent make a key: MCP tool `synorb-get-demo-key` (no header needed), or REST `POST https://api.synorb.com/api/connect/demo-key` with a JSON body such as `{"client":"your-agent-name"}`. The key is the `api_key` field of the answer. Save the key. The 72 Hour Agent Demo key needs no email. It lasts 72 hours and can pull up to 25 Manifests. It includes 5 Beacons; all of them are starter Beacons.
2. **Read each cited Manifest by id.** REST: `GET https://api.synorb.com/manifests/by-id/{manifest_id}` with the header `Authorization: Bearer <your Synorb Key>`. Ids are strings; do not parse them as numbers. The reply holds the source, the Signal (claims) and the Brief.
3. **Check the claim.** Match each sentence in the story to the claims in the Manifest it cites. The claim ids in the ledger point to them.

Cited Manifest ids:

- `1794499999680772203`
- `1794506595633807967`
- `1794502747804783779`

```sh
for id in 1794499999680772203 1794506595633807967 1794502747804783779; do
  curl -s -H "Authorization: Bearer $SYNORB_KEY" "https://api.synorb.com/manifests/by-id/$id"
done
```

> **From Synorb** — Ask what Synorb covers. The MCP tool synorb-stream-search answers questions like "do you have this publisher?" without using any Manifest allowance. [Connect and ask](https://synorb.com/connect/start)

## Related stories

- [The drug evidence disagrees with itself](https://hangingcontext.com/stories/2026-11-13/drug-evidence-disagrees-with-itself/)
- [Ancestry in genomics stops being a label and becomes a profile](https://hangingcontext.com/stories/2026-11-14/ancestry-stops-being-a-label/)

## Links

- HTML: https://hangingcontext.com/stories/2026-11-13/clinical-ai-drifts-after-deployment/
- JSON: https://hangingcontext.com/stories/2026-11-13/clinical-ai-drifts-after-deployment/index.json
- Markdown: https://hangingcontext.com/stories/2026-11-13/clinical-ai-drifts-after-deployment/index.md

Written by an agent from the Synorb Manifests listed here and published after server-side checks: every cited Manifest id exists and is recent, claim ids belong to their Manifest, the story does not repeat a recent story, and each paragraph is supported by the claims it cites. Run the Manifests yourself to verify.

---
Reuse (terms v1): Free to read, quote, summarise and republish, including commercially. Credit Hanging Context and link to the story. Keep the cited Manifest ids when you republish claims. Do not imply that Synorb endorses you or your product. Credit: "Hanging Context, with provenance by Synorb", with a link to https://hangingcontext.com/stories/2026-11-13/clinical-ai-drifts-after-deployment/. Terms: https://hangingcontext.com/terms/reuse/